Categories
Product Details
Overview
The newest high-performance Scopemeter model with widened bandwidth and key control on each channel. The device is convenient for signals research, operations between the channels and the electrical parameters measurement as well.
Classic Scopemeter (Digital storage oscilloscope + Digital Multimeter). DSO: Two channels; bandwidth 100 MHz; sample rate (Real time) 500MS/s; record length 6K points on each channel; cursor and automeasurement (20 types); trigger type: edge, video; Math; FFT; peak detect and average; waveform record; USB Flash Disk Storage. DMM: VDC, VAC, DC, AC, resistance, capacitance, diode test, on-off test.. Display: 3.8 '' color LCD, TFT screen, 640×480 pixels. Dimensions: 7.1in x 4.5in x 1.6in / 180mm x 115mm x 40mm. Weight: 22.8oz / 645g
Technical Specifications:
The instrument should operate continuously for more than 30 minutes under the specified operating temperature.
If the operating temperature is up to or larger than 41°F / 5°C, the system function menu must be opened to make the system perform a “Auto- calibration” procedure.
Except those specifications marked with the word Typical, all specifications can be up to the mentioned definitions.
Sampling
- Sampling modes: Normal sampling, Peak detection, Average value
- Sampling rate: 500 MSa/s
Input
- Input coupling: DC, AC, Ground
- Input impedance: 1M Ω±2% connected in parallel with 15pF ± 5pF
- Probe attenuation coefficient: 1X, 10X, 100X, 1000X
- Max. Input voltage: 400V (peak)
- Channel delay time (typical): 150ps
Horizontal
- Sampling rate range: 0.25S/s~500MS/s
- Waveform interpolation: (sin x)/x
- Record length: 6K points on each channel
- Scanning speed range (S/div): 5ns/div~100s/div, stepping in the “1-2-5” mode
- Sampling rate and delay time accuracy: ±100ppm (any time interval which is equal to or larger than 1ms)
- Time interval (ΔT) measurement accuracy (full bandwidth): Single: ± (1 sampling interval time +100ppm × reading+0.6ns) > average 16 : ± (1 sampling interval time +100ppm × reading+0.4ns)
Vertical
- Analog digital converter (A/D): With the resolution of 8 bits, make sampling on both channels synchronously
- Sensitivity range (V/div): 5mV/div~5V/div (at the input BNC)
- Displacement range: ±1V(5mV/div~100mV/div), ±10V(200mV/div~1V /div), ±50V(2V/div~5V /div)
- Analog bandwidth: 100MHz
- Single bandwidth: Full bandwidth
- Low frequency response (AD coupling, -3dB): ≥5Hz (at the BNC)
- Rise time (typical one at the BNC): ≤3.5ns
- DC gain accuracy: ±3%
- DC measurement accuracy (average value sampling mode): The voltage difference (ΔV) between any two points on the waveform after averaging the captured waveforms more than 16: ±(5% reading + 0.05 divisions).
Trigger sensitivitytriggering
(Edge)
- DC coupling 1div (DC~full bandwidth)
- AC coupling Same as the DC coupling when it is equal to or larger than 50Hz
- Triggering level range ±6 divisions from the screen center
- Triggering level accuracy (typical) which is applicable to the signal with rise and fall time equal to or longer than 20ns ±0.3 divisions
- Trigger displacement 655 divisions for pre-triggering and 4 divisions for post-triggering
- Trigger Holdoff range 100ns~10s
- Make a 50% level setting (Typical). Operation with the input signal frequency equal to or larger than 50Hz
- Trigger sensitivity (Video triggering and typical mode) 2 divisions of peak-to-peak value
- Signal system and line/field frequency (Video triggering mode); Support the NTSC, PAL and SECAM broadcasting systems of any field or line frequency.
Cymometer
- Readout resolution: 6 digits
- Frequence range: AC coupling, 2Hz to full bandwidth
- Single source: when the triggering mode is edge triggering it is a one channel cymometer; When the triggering mode is alternating triggering it is a two channel cymometer; When the triggering mode is video triggering the cymometer will not work
- Cursor measurement: Voltage difference (ΔV) and time difference (ΔT) between cursors
- Auto measurement: Peak-to-peak value, average value, root mean square value, frequency, period, Vmax, Vmin, Vtop, Vbase, Vamp, Overshoot, Preshoot, RiseTime, Fall Time, +Width, -Width, +Duty, -Duty, Delay A B.
Probe
1X position | 10X position | |
---|---|---|
Bandwidth | Up to 6 MHz (DC) | Up to full bandwidth (DC) |
Attenuation rate | 1: 1 | 10: 1 |
Compensation range | 15pf~35pf | |
Input resistance | 1MΩ±2% | 10MΩ±2% |
Input impedance | 85pf~115pf | 14.5pf~17.5pf |
Input voltage | <200 V DC+Peak AC | <600 V DC+Peak AC |
Voltage (VDC)
Input Impedance: 10MΩ. Max. Input Voltage: 1000V (DC or AC peak-to-peak value)
Range | Accuracy | Resolution |
---|---|---|
400.0mv | 100µV | |
4.000V | 1mV | |
40.00V | ±1%±2 digit | 10mV |
400.0V | 100mV | |
1000.0V | 1V |
Voltage (VAC)
Input Impedance: 10MΩ. Max. Input Voltage: 750V (AC, virtual value). Frequency range: from 40Hz to 400Hz. Display: Virtual value of the sine wave
Range | Accuracy | Resolution |
---|---|---|
4.000V | 1mV | |
40.00V | ±1%±3 digits | 10mV |
400.0V | 100mV | |
750.0V | ±1.5%±3 digit | 1V |
Direct Current (DC)
Range | Accuracy | Resolution |
---|---|---|
40.00mA | ±1.5%±1 digit | 10µA |
400.0mA | ±1.5%±1 digit | 100µA |
10A | ±3%±3 digit | 10mA |
Alternating Current (AC)
Range | Accuracy | Resolution |
---|---|---|
40.00mA | ±1.5%±3 digit | 10µA |
400.0mA | ±2%±1 digit | 100µA |
10A | ±5%±3digit | 10mA |
Resistance
Range | Accuracy | Resolution |
---|---|---|
400.0Ω | ±1%±3 digit | 0.1Ω |
4.000KΩ | 1Ω | |
40.00KΩ | 10Ω | |
400.0KΩ | ±1%±1 digit | 100Ω |
4.000MΩ | 1KΩ | |
40.000MΩ | ±1,5%±3 digit | 10KΩ |
Capacitance
Range | Accuracy | Resolution |
---|---|---|
51.20nF | 10pF | |
512.0nF | 100pF | |
5.120uF | ±3%±3 digit | 1nF |
51.20uF | 10nF | |
100uF | 100nF |
Diode
Voltage reading: 0 V~1.5 V
On-off Test
There is a beep sound when the on-resistance is less than 50Ω.
General Specifications:
Basic parameters
- Mechanical dimensions: 7.1in x 4.5in x 1.6in / 180mm x 115mm x 40mm
- Weight: 22.8oz / 645g
- Power consumption: < 6 W
- Display type: 3.7" color liquid crystal display
- Display resolution: 640 (horizontal) × 480 (vertical) pixels
- Display color: 65536 colors
Power Adapter
- Power supply: 100-240 V AC 50/60Hz
- Power output: 9 VDC
- Current output: 3000mA
Battery type: 7.4V built-in Li-ion battery
Working environment
- Used battery: 32 to 122 °F / 0 to 50 °C
- Power adapter: 32 to 104 °F / 0 to 40 °C
Temperature:
Operation
- 32 to 50 °F / 0 to 10°C, no condensation
- 50 to 86 °F / 10 to 30°C, 95%
- 86 to 104 °F / 30 to 40°C, 75%
- 104 to 122 °F / 40 to 50 °C, 45 %
Storage
- -4 to +140 °F / -20 to +60 °C, no condensation